Giving Shape to Ideas

search form

Specifications

Measuring Instruments

Spectroradiometer CS-2000

NEW

High precision spectral radiance/chromaticity measurement possible from 0.003 cd/m²

Specifications

Principal Specifications CS-2000

Model CS-2000
Wavelength range 380 to 780 nm
Wavelength resolution 0.9 nm/pixel
Display wavelength bandwidth 1.0 nm
Wavelength precision ±0.3 nm (median wavelength: 435.8 nm, 546.1 nm, 643.8 nm Hg-Cd lamp)
Spectral bandwidth 5 nm or less (half bandwidth)
Measuring angle (selectable) 0.2° 0.1°
Measurement luminance range (Illuminant A) 0.003 to 5,000 cd/m2 0.075 to 125,000 cd/m2 0.3 to 500,000 cd/m2
Minimum luminance display 0.00002 cd/m2
Accuracy: Luminance (Illuminant A) *1 ±2%
Accuracy: Chromaticity (Illuminant A)*1 x,y :±0.003 (0.003 to 0.005 cd/m2)
x,y :±0.002 (0.005 to 0.05 cd/m2)
x :±0.0015 (0.05 cd/m2 to)
y :±0.001
x,y :±0.003 (0.075 to 0.125 cd/m2)
x,y :±0.002 (0.125 to 1.25 cd/m2)
x :±0.0015 (1.25 cd/m2 to )
y :±0.001
x,y :±0.003 (0.3 to 0.5 cd/m2)
x,y :±0.002 (0.5 to 5 cd/m2)
x :±0.0015 (5 cd/m2 to )
y :±0.001
Repeatability: Luminance(2σ) 0.4 % (0.003 to 0.05 cd/m2)
0.3 % (0.05 to 0.1 cd/m2)
0.15 % (0.1 to 5,000 cd/m2)
0.4 % (0.075 to 1.25 cd/m2)
0.3 % (1.25 to 2.5 cd/m2)
0.15 % (2.5 to 125,000 cd/m2)
0.4 % (0.3 to 5 cd/m2)
0.3 % (5 to 10 cd/m2)
0.15 % (10 to 500,000 cd/m2)
Repeatability: Chromaticity(2σ) 0.002 (0.003 to 0.005 cd/m2)
0.001 (0.005 to 0.1 cd/m2)
0.0006 (0.1 to 0.2 cd/m2)
0.0004 (0.2 to 5,000 cd/m2)
0.002 (0.075 to 0.125 cd/m2)
0.001 (0.125 to 2.5 cd/m2)
0.0006 (2.5 to 5 cd/m2)
0.0004 (5 to 125,000 cd/m2)
0.002 (0.3 to 0.5 cd/m2)
0.001 (0.5 to 10 cd/m2)
0.0006 (10 to 20 cd/m2)
0.0004 (20 to 500,000 cd/m2)
Polarization error 1°: 2% or less (400 to 780 nm); 0.1° and 0.2°: 3% or less (400 to 780 nm)
Integration time Fast: 0.005 to 16 sec; Normal: 0.005 to 120 sec.
Measurement time 2 sec min. (Manual mode) to 243 sec. max. (Normal mode)
Color space mode LVxy, Lvu'v', LvTΔuv, XYZ, dominant wavelength, spectral graph
Interface USB 1.1
Operating temperature/humidity range 5 to 35°C, relative humidity 80% or less with no condensation
Storage temperature/humidity range 0 to 45°C, relative humidity 80% or less with no condensation
Size Main unit: 158 (W) × 200 (H) × 300 (D) mm; Lens: Φ70 × 95 mm
Weight 6.2 kg
*1
Average of 10 measurements in Normal mode at a temperature of 23±2°C and a relative humidity of 65% or less.
*
Specifications are subject to change without prior notice.

Data Management Software

CS-S10w Professional (Standard accessory)

With this software, the CS-2000 can be controlled from a personal computer to display measured data with various graphs or lists, to transfer data to spreadsheet software, or to copy-and-paste data. The CS-S10w offers various data management, analysis and evaluation options, which help research and development or quality control.

Ugrás a tetejére