Measuring Instruments

NEW
High precision spectral radiance/chromaticity measurement possible from 0.003 cd/m²
| Model | CS-2000 | ||
|---|---|---|---|
| Wavelength range | 380 to 780 nm | ||
| Wavelength resolution | 0.9 nm/pixel | ||
| Display wavelength bandwidth | 1.0 nm | ||
| Wavelength precision | ±0.3 nm (median wavelength: 435.8 nm, 546.1 nm, 643.8 nm Hg-Cd lamp) | ||
| Spectral bandwidth | 5 nm or less (half bandwidth) | ||
| Measuring angle (selectable) | 1° | 0.2° | 0.1° |
| Measurement luminance range (Illuminant A) | 0.003 to 5,000 cd/m2 | 0.075 to 125,000 cd/m2 | 0.3 to 500,000 cd/m2 |
| Minimum luminance display | 0.00002 cd/m2 | ||
| Accuracy: Luminance (Illuminant A) *1 | ±2% | ||
| Accuracy: Chromaticity (Illuminant A)*1 | x,y :±0.003 (0.003 to 0.005 cd/m2) x,y :±0.002 (0.005 to 0.05 cd/m2) x :±0.0015 (0.05 cd/m2 to) y :±0.001 |
x,y :±0.003 (0.075 to 0.125 cd/m2) x,y :±0.002 (0.125 to 1.25 cd/m2) x :±0.0015 (1.25 cd/m2 to ) y :±0.001 |
x,y :±0.003 (0.3 to 0.5 cd/m2) x,y :±0.002 (0.5 to 5 cd/m2) x :±0.0015 (5 cd/m2 to ) y :±0.001 |
| Repeatability: Luminance(2σ) | 0.4 % (0.003 to 0.05 cd/m2) 0.3 % (0.05 to 0.1 cd/m2) 0.15 % (0.1 to 5,000 cd/m2) |
0.4 % (0.075 to 1.25 cd/m2) 0.3 % (1.25 to 2.5 cd/m2) 0.15 % (2.5 to 125,000 cd/m2) |
0.4 % (0.3 to 5 cd/m2) 0.3 % (5 to 10 cd/m2) 0.15 % (10 to 500,000 cd/m2) |
| Repeatability: Chromaticity(2σ) | 0.002 (0.003 to 0.005 cd/m2) 0.001 (0.005 to 0.1 cd/m2) 0.0006 (0.1 to 0.2 cd/m2) 0.0004 (0.2 to 5,000 cd/m2) |
0.002 (0.075 to 0.125 cd/m2) 0.001 (0.125 to 2.5 cd/m2) 0.0006 (2.5 to 5 cd/m2) 0.0004 (5 to 125,000 cd/m2) |
0.002 (0.3 to 0.5 cd/m2) 0.001 (0.5 to 10 cd/m2) 0.0006 (10 to 20 cd/m2) 0.0004 (20 to 500,000 cd/m2) |
| Polarization error | 1°: 2% or less (400 to 780 nm); 0.1° and 0.2°: 3% or less (400 to 780 nm) | ||
| Integration time | Fast: 0.005 to 16 sec; Normal: 0.005 to 120 sec. | ||
| Measurement time | 2 sec min. (Manual mode) to 243 sec. max. (Normal mode) | ||
| Color space mode | LVxy, Lvu'v', LvTΔuv, XYZ, dominant wavelength, spectral graph | ||
| Interface | USB 1.1 | ||
| Operating temperature/humidity range | 5 to 35°C, relative humidity 80% or less with no condensation | ||
| Storage temperature/humidity range | 0 to 45°C, relative humidity 80% or less with no condensation | ||
| Size | Main unit: 158 (W) × 200 (H) × 300 (D) mm; Lens: Φ70 × 95 mm | ||
| Weight | 6.2 kg | ||
With this software, the CS-2000 can be controlled from a personal computer to display measured data with various graphs or lists, to transfer data to spreadsheet software, or to copy-and-paste data. The CS-S10w offers various data management, analysis and evaluation options, which help research and development or quality control.